Overview
--------

PyChop is a tool to allow direct inelastic neutron scattering users to estimate
the inelastic resolution and incident flux for a given spectrometer setting.
Currently, the four direct geometry spectrometers at ISIS (LET, MAPS, MARI, and
MERLIN) are supported. 

For MERLIN and LET, in addition, PyChop will also calculate the allowed Ei's in
multi-rep mode, and plot the time-distance diagrams for the desired setting.

Options
-------

First, the instrument, chopper slit packages (or instrument configurations for
LET) and chopper frequency(ies) have to be selected from the pull-down menus 
(combo boxes). Then the user should enter the desired incident energy (or the Ei
to focus on for multi-rep operation) in the line edit box. 

Clicking the *Calculate* button will cause PyChop to run the resolution and flux
calculations, which will take 1-2s and update the plots in the tabs on the right
hand side of the GUI. Alternatively, in the options menu, the user can select
the option of having the calculations run when enter (return) is pressed when
the focus is on the Ei line edit box. 

If the *Hold current plot* check box is enabled (selected) then the resolution
vs energy plots will overplot the current axes. 

If the *Show multi-reps* check box is enabled (selected) then the resolution vs
energy for all allowed Ei's will be plotted on the current axes. This only
applies to LET or MERLIN with the G (gadolinium) chopper slit package.

In the *Flux-Ei* plot tab, there is a slider at the bottom and a line edit box 
to allow the user to select the maximum x-range (incident energy range) to plot.
The plot updates when the slider or edit box is changed. To save computation 
time, the flux / elastic resolution is only calculated at twenty incident energy
points from 0.1 meV to the maximum selected.

The flux and elastic resolution as a function of chopper frequency for the
specified Ei is shown in the *Flux-Freq* plot tab. If *Hold current plot* is
selected then several settings can be overplotted. The program will not overplot
if it detects that only the frequency has changed.

If the instrument is LET (or MERLIN with the G chopper), the time-distance plot 
is enabled, and an additional option to change the phase of chopper 2 is 
luz.paz's avatar
luz.paz committed
available. This chopper has a wide opening and can be used to suppress low 
energy reps. The time delay which is specified in the chopper 2 phase edit box 
is the time-of-flight in microseconds relative to the moderator pulse when the 
chopper first opens.

If the *Instrument scientist mode* option is selected, a similar option is
enabled for MERLIN if the G chopper is used. In this case, the phase (time
delay) of the thick disk chopper can be adjusted. The time delay is the time-of-
flight at which the chopper slit first opens (sweeps across the beam profile).

The Matplotlib axes showing the calculated data have the standard toolbars.

Command line interface
----------------------

In addition to the GUI, there is also a python commandline interface to PyChop.
This is encapsulated in the ``PyChop2`` class within the ``PyChop`` module. Within
MantidPlot, to do a single point calculation of the flux and resolution

.. code:: python 

    from PyChop import PyChop2
    resolution, flux = PyChop2.calculate(inst='maps', chtyp='a', freq=500, ei=600, etrans=range(0,550,50))

The parameters are in order, so ``PyChop2.calculate('maps','a',500,600,range(0,550,50))``
also works.

In addition, an object orient interface is provided:

.. code:: python 

    mapsres = PyChop2('maps')
    mapsres.setChopper('a')
    mapsres.setFrequency(500)
    mapsres.setEi(600)
    res = mapsres.getResolution(range(0,550,50))

In particular, the method ``getResolution``, which takes the energy transfers to
calculate the resolution for as an input, can be directly passed to third party 
programs for resolution convolution purposes.

For further help, use ``help(PyChop2)`` after importing the class.

Theory
------

The energy resolution calculated by ``PyChop`` has contributions from the time
width of the moderator pulse :math:`\tau_{\mathrm{mod}}`, the opening times of the 
choppers, :math:`\tau_{\mathrm{chop}}`, the response time of the detector,
:math:`\tau_{\mathrm{det}}`, and the effect of the sample, :math:`\tau_{\mathrm{sam}}`.
The first two contributions dominate so we will only concentrate on those.

The moderator time width is determined from fitting data above 100 meV to a 
Duc Le's avatar
Duc Le committed
:math:`\chi^2` distribution `[1]`_ which has a variance :math:`\tau_{\mathrm{mod}}^2
=3/(\Sigma v)^2` where :math:`\Sigma` is the macroscopic scattering cross-section
of the moderator and :math:`v` is the neutron velocity. However, experimentally
Duc Le's avatar
Duc Le committed
it was found that this underestimates the widths at high energy `[2]`_, so that a 
modified form for the variance

Duc Le's avatar
Duc Le committed
.. math:: \tau_{\mathrm{mod}}^2 = \tau_0 + \frac{3}{(\Sigma v)^2}

is used in PyChop. In future versions, the moderator lineshape will be reparameterised
to use an Ikeda-Carpenter lineshape, which more accurately describes the ToF spectrum
at lower neutron energies.

The chopper time width is determined from the geometry of chopper and is given by
Duc Le's avatar
Duc Le committed
`[2]`_, `[3]`_

.. math:: 
        \tau_{\mathrm{chop}}^2 \left\{ \begin{array}{ll} \frac{(\Delta T)^2}{6} 
        \left[\frac{1-\gamma^4/10}{1-\gamma^2/6}\right] & 0 \leq \gamma < 1 \\
        \frac{(\Delta T)^2}{6} \left[\frac{3}{5} 
        \frac{\gamma(\sqrt{\gamma}-2)^2(\sqrt{\gamma}+8)}{\sqrt{\gamma}+4}\right]
        & 1 \leq \gamma < 4 \\
        \mathrm{undefined} & \gamma \geq 4 \end{array} \right.

where

.. math:: \begin{array}{rcl} \Delta T &=& \frac{p}{2R\omega} \\
        \gamma &=& \frac{2R}{\Delta T} \left| \frac{1}{s} - \frac{1}{v} \right| \\
        s &=& 2\omega\rho \end{array}

and :math:`p` is the width of the slits of the Fermi chopper, :math:`R` is the radius
of the chopper package (assumed cylindrical), :math:`\omega` is its rotation speed, 
:math:`v` is the neutron velocity and :math:`\rho` is the curvature of Fermi chopper 
slits.

The time variances above are defined at the moderator and chopper positions respectively.
As the neutron bunches travel towards the sample and detector they also spread out,
and the final time (energy) widths are determined by the geometry (distances) of the
instrument. Specifically, the relative energy width is given by the sum in quadrature
of each of the contributing time widths, which we will restrict here to the two major
Duc Le's avatar
Duc Le committed
terms, :math:`\tau_{\mathrm{mod}}` and :math:`\tau_{\mathrm{chop}}` `[4]`_:

.. math:: \left( \frac{\Delta E}{E_i}\right )^2 = 
        \left[ 2\frac{\tau_{\mathrm{chop}}}{t_{\mathrm{chop}}} \left(1+\frac{l_0+l_1}{l_2}
        \left(\frac{E_f}{E_i}\right)^{\frac{3}{2}} \right) \right]^2
        + \left[ 2\frac{\tau_{\mathrm{mod}}}{t_{\mathrm{chop}}} \left(1+\frac{l_1}{l_2}
        \left(\frac{E_f}{E_i}\right)^{\frac{3}{2}} \right) \right]^2

where :math:`t_{\mathrm{chop}}` is the time of arrival of the neutron bunch at the
Fermi (or final resolution disk) chopper, :math:`l_0` is the moderator-chopper,
:math:`l_1` the chopper-sample and :math:`l_2` the sample-detector distance. :math:`E_i`
and :math:`E_f` are the incident and scattered neutron energies.

The flux is obtained from lookup tables of measured (white-beam) flux on each instrument.


Sample contribution
-------------------

Although the contribution of a sample to the resolution of direct-geometry chopper spectrometers
is usually negligible,
it is currently included by calculating of the variance of the time-of-flight due to the size
of the sample.
So far only two shapes are supported: plate and thin annulus.
The broadening caused by a plate sample is calculated as proportional
to :math:`\frac{1}{12} w^2`, where :math:`w` is the width of the plate.
The factor :math:`\frac{1}{12}` comes from the variance of a uniform distribution

.. math:: \int^{\frac{1}{2}}_{-\frac{1}{2}} x^2 dx = \frac{1}{12}

.. figure:: /images/Pychop-annulus-shape.png
   :alt: sample-annulus-variation.png
   :align: right
   :width: 300

For a thin annulus, the variation is proportional to its diameter.
The fractional factor is calculated as the following variance

.. math:: \frac{ \int^{\frac{1}{2}}_{-\frac{1}{2}} x^2 \rho(x) dx } {\int \rho(x) dx} 
	  
Here :math:`x= r \cos\theta = \frac{1}{2} \cos\theta`.
:math:`\rho(x)` is the (unnormalized) distribution function,
which is proportional :math:`\frac{1}{\sin\theta}`.
So the integration evaluates to

.. math:: \frac{ \int_{0}^{\pi} x^2 d\theta } {\int d\theta} = \frac{1}{8}

Duc Le's avatar
Duc Le committed
References
----------
Duc Le's avatar
Duc Le committed
.. _[1]:
Duc Le's avatar
Duc Le committed
[1] `RAL-94-025: The resolution function of the chopper spectrometer HET at ISIS,
T G Perring, Proceedings of ICANS XII (1993)
<http://www.neutronresearch.com/parch/1993/01/199301013280.pdf>`_
Duc Le's avatar
Duc Le committed
.. _[2]: 

[2] RALT-028-94: High energy magnetic excitations in hexagonal cobalt,
T G Perring, Ph.D. Thesis, University of Cambridge (1991)

.. _[3]:

[3] `M. Marseguerra and G. Pauli, Neutron transmission probability through a 
curved revolving slit, Nucl. Inst. Meth. 4 (1959) 140
<http://dx.doi.org/10.1016/0029-554X(59)90066-7>`_

.. _[4]:

[4] RAL-85-052: MARS - A Multi-Angle Rotor Spectrometer for the SNS,
C J Carlile, A D Taylor and W G Williams (1985)

.. categories:: Interfaces